Dual Combustion Furnace Infrared Carbon and Sulfur Analyzer CS-3500G which can be used to detect the content of carbon or sulfur in metals, alloys and organic materials.
CS 3500G Dual Combustion Furnace Infrared Carbon and Sulfur Analyzer, is a high-end combined infrared carbon and sulfur analyzer of world-leading technology.
Due to the diversity of specimens of users and analytical methods, the instrument is equipped with two sets of sample pre-treatment units, i.e. high frequency induction furnace and tubular resistance furnace. The resistance furnace and the high frequency induction furnace share the same infrared detection system for the detection of content of output gases to be detected. It is easy to switch between the two furnaces.
CS 3500G Dual Combustion Furnace Infrared Carbon and Sulfur Analyzer can be used for the detection of carbon and sulfur in a wide range of specimens, such as steel, cast iron, alloy, metal ore, ceramics, cement, lime, rubber, coal,coke, refractory, carbide, graphite, oil, catalyst, soil and other solid materials.
Scope of analysis:
High frequency induction furnace (500mg sample): low carbon: 1ppm ~ 0.1%, high carbon: 0.1% ~ 5%
Low sulfur: 1ppm ~ 0.3% high sulfur (150mg sample): 0.3% ~ 30%
Tubular resistance furnace (500mg sample): carbon: 0.01% ~ 20%
Sulfur: 0.005% ~ 20%*
Note: The number and size of measuring cell can be selected according to the sample content requirements of users.
The analysis range can be extended to 100% by changing the sample weight.
Analysis accuracy:
High frequency induction furnace: low carbon: 1ppm or RSD ≤ 1%, high carbon: RSD ≤ 0.5%
Low sulfur: 1.5ppm or RSD ≤ 1.5%, high sulfur: RSD ≤ 1.5%
Tubular resistance furnace: low carbon: RSD ≤ 1%, high carbon: RSD ≤ 0.5%
Low sulfur: RSD ≤ 1.5%, high sulfur: RSD ≤ 1.5%
Note: The standard deviation of standard sample shall prevail.
Sensitivity: 0.01ppm
Analysis time:
High frequency induction furnace: 40-60 seconds
Tubular resistance furnace: 150s ~ 300s
Sample weighing:
High frequency induction furnace: 0.1g-0.5g;
Tubular resistance furnace: 300mg / coal (typical value)
Note: The computer is connected with the electronic balance to realize accurate weighing, and the sample quality is automatically transmitted to the analysis software. If necessary, the sample quality can also be entered manually.
Combustion furnace:
High frequency induction furnace: the combustion chamber of 18mhz 2.7kva high frequency furnace can be automatically cleaned in each analysis process, which is easy for daily maintenance.
Tubular resistance furnace: max.20a, max.1550 ℃, temperature continuously adjustable.
Crucible pretreatment
Ceramic crucible for high frequency induction furnace: diameter: 25 mm, height: 27 mm
Tubular resistance furnace: 87mm in length, 12mm in width and 10mm in height
Power supply
Analyzer power supply 220 VAC ± 10% 50 ± Hz 16A
Resistance furnace power supply: 220 VAC ± 10% 50 ± Hz 20A
SIZE
Tube resistance furnace : 33cm wide, 52cm high and 60cm thick
High frequency induction furnace is 55cm wide, 77cm high and 76cm thick
Weight
High frequency induction furnace 100kg;
Tube resistance furnace 30kg
Germany produced solid-state infrared detector components
Flexible control of the heating power
Support for full manual mode
Advanced infrared thermostatic technology
Advanced flow control technology
Unique computer software
First-class linear treatment effect
High testing precision
Highly reliable circuit components; rich self-diagnostic functions
Multiple circuit protections against meltdown
Full close-down gas channel with zero death volume
The application of infrared absorption following high frequency combustion or following tube furnace combustion
enables quick analysis of the content of carbon and sulfur in solid inorganics.
Software Functions:
1) Detection of the movement of solenoid valves on gas channels
2) Detection and adjustment of infrared signals
3) Fast display of analytical results and analytic curves
4) Automatic store of analytical results
5) Drawing and store of dynamic release curves
6) Specimen weight value entry: the software can automatically read weight values of specimens
7) Channel Switch: automatic switch among high-carbon channel, low-carbon channel and sulfur channel
Data Processing:
1) Data stored can be referred in multiple ways, such as by date or specimen ID
2) Screening of data
3) The statistic treatment of data (mean, standard deviation, relative deviation)
4)Signal comparison to the release curves of the analytical results