Carbon & Sulfur Analyzer

CS-3000

CS-3000 Series Analyzer is ideal to provide pricise and quick determination of carbon and sulfur simultaneously in primary steels, ores, finished metals, ceramics, and other inorganic materials.

A sample of typically 0.5 to 1.0g is weight into a crucible together with 1.5g of tungsten accelerator. This crucible is placed on a pedestal which drives it into an induction furnace by a pneumatic system. The sample is completely burned in the furnace by pure Oxygen.The Carbon of this sample is oxidized to CO2 and the Sulphur content to SO2. Dust and other particles are trapped in a filter system.The CO2 and SO2 is transported from the furnace in the Oxygen carrier gas to the infrared adsorption detectors.The actual gas concentrations are determined and the data are send to the PC.The operation software is calculating the results after integration of the whole signal peaks. A calibration with known samples builds up a database which created the necessary accuracy and precision. The burned crucible is driven out of the furnace by the pneumatic system and disposed by an operator. The furnace and gas flow system is purged after analysis by clean Oxygen to avoid any interference (memory effect) of the individual samples.

DescriptionFeaturesParametersBrochure

A sample of typically 0.5 to 1.0g is weight into a crucible together with 1.5g of tungsten accelerator. This crucible is placed on a pedestal which drives it into an induction furnace by a pneumatic system. The sample is completely burned in the furnace by pure Oxygen.The Carbon of this sample is oxidized to CO2 and the Sulphur content to SO2. Dust and other particles are trapped in a filter system.The CO2 and SO2 is transported from the furnace in the Oxygen carrier gas to the infrared adsorption detectors.The actual gas concentrations are determined and the data are send to the PC.The operation software is calculating the results after integration of the whole signal peaks. A calibration with known samples builds up a database which created the necessary accuracy and precision. The burned crucible is driven out of the furnace by the pneumatic system and disposed by an operator. The furnace and gas flow system is purged after analysis by clean Oxygen to avoid any interference (memory effect) of the individual samples.

The induction furnace ensures a fast and complete combustion at temperatures of about 2200°C. It’s robust design ensures a long lifetime and minimized maintenance. With 18.5 MHz and 2.2 kVA its a suitable furnace for the combustion of a wide range of materials.

The unique design reduces a splattern and the dust from the combustion is trapped in an easy to clean metal filter.

The specialty for this application developed IR detector with up to 4 independent absorption units ensures the needed wide analytical range with high precision.The reliable design is known for long term stability and very low maintenance. 

High tech detectors and interference filters are the base for a long time stability of the calibration. A build in temperature controller makes the detector system independent from the environmental conditions.

Solid-state pyroelectric detector made in Germany
Flexible heating power control
Advanced temperature control technology
Advanced gas flow control technology
Software has unique features
IR signal data processing result is ace after linearizing
Analytical precision reached international advance level
Electric parts with high reliability
Rich in diagnosis function
Redundancy protection against meltdown
Sealed gas passage with no dead volume

Scope of analysis:

LC: 0.0001%~0.2%

HC: 0.2%~6%

LS: 0.0001%~0.3%

HS: 0.3%~30%

Note:*The quantity and size of measuring cells can be determined under requirements for the content of user’s sample.

 Analysis precision:

Carbon: LC: 1ppm or RSD≤1%

HC: RSD≤0.5%     

Sulfur: LS: 1.5ppm or RSD≤1.5%

HS: RSD≤1.5%

Note: *The precision is subject to values not larger than standard deviation of standard samples.

Sensitivity:

C and S: 0.01ppm

Time for analysis:

        30s~40s

Sample weight:

        0.2g~0.5g

Power supply:

220VAC ±10% 50/60Hz16A

Size:

W 55cm  H 77cm  D 76cm

Weight:

100kg

Carrier gas:
        Oxygen: purity ≥99.5%, pressure 0.30 MPa, pressure of furnace end 0.05MPa, 3L/min
Pneumatic gas:
        Nitrogen or compressed air (purified); pressure imported 0.30 MPa